Try the questions in the test above first, then open each one here to see the correct answer and why every other option is wrong.
01A technician checks an ECG monitor against the WHO maintenance procedure, which requires heart rate meter accuracy within ±3% at three points. With an arrhythmia simulator set to 80 bpm, the monitor displays 83 bpm. What is the correct conclusion?
Practice question Unit: Biomedical Equipment Technology
APass, because the error is only 3 bpm, which is a small absolute value
The tolerance is relative, so a small absolute difference is not enough when it exceeds 3% of the setting.
BPass, because the error is about 2.5% of the displayed value
The error should be taken relative to the simulator setting of 80 bpm, which gives 3.75%, not 2.5%.
CFail, because the error is 3.75%, which is outside the allowed limit ✓ Correct answer
The error is 3 bpm out of 80, which is 3.75%, and that exceeds the 3% limit given in the procedure.
DFail, because the error is about 7.5% and the monitor must be replaced
A 3 bpm difference on 80 bpm is 3.75%, not 7.5%; the unit needs adjustment or repair, not automatic replacement.
Reference: WHO, Medical equipment maintenance programme overview (WHO Medical device technical series, 2011), Appendix B.4 Monitor, ECG, p. 66; SCFHS Medical Devices – Technician Content Guideline (V1.220901), Biomedical Equipment Technology, p. 2
Ask a friend
02A 12.6 V battery supplies a 4.0 A load, and its terminals then read 11.8 V. Assuming the internal resistance stays constant, what current would flow if the terminals were shorted through negligible resistance?
Practice question Unit: Advanced Electronics and Electricity
A4.3 A
This is the emf divided by the load resistance only, which ignores the short-circuit condition entirely.
B15.8 A
This divides emf by the 0.8 V drop, treating a voltage as if it were the internal resistance.
C59 A
This divides the terminal voltage, not the emf, by the internal resistance.
D63 A ✓ Correct answer
The internal drop is 12.6 − 11.8 = 0.8 V at 4.0 A, so r = 0.2 Ω, and shorted current is 12.6/0.2 = 63 A.
Reference: OpenStax, College Physics 2e, Section 21.2 Electromotive Force: Terminal Voltage; SCFHS Medical Devices – Technician Content Guideline (V1.220901), Advanced Electronics and Electricity, p. 2
Ask a friend
03A resting echocardiogram records an end-diastolic volume of 130 mL and an end-systolic volume of 60 mL. The heart rate is 80 beats per minute. What is the cardiac output?
Practice question Unit: Anatomy and Physiology
A0.88 L/min
This divides stroke volume by heart rate; cardiac output is the product of the two, not their quotient.
B4.8 L/min
This multiplies the heart rate by the end-systolic volume of 60 mL instead of the stroke volume.
C5.6 L/min ✓ Correct answer
Stroke volume is 130 − 60 = 70 mL, and 70 mL × 80 beats per minute gives 5.6 L/min.
D10.4 L/min
This multiplies the heart rate by the end-diastolic volume of 130 mL, which is not the blood actually ejected.
Reference: OpenStax, Anatomy and Physiology 2e, Section 19.4 Cardiac Physiology; SCFHS Medical Devices – Technician Content Guideline (V1.220901), Anatomy and Physiology, p. 2
Ask a friend
04An amplifier raises a signal level by 30 dB (power). Using dB = 10 log(P2/P1), by what factor does the power increase?
Practice question Unit: Mathematics
A3
Dividing 30 by 10 gives the exponent, not the factor; the factor is 10 raised to it.
B31.6
This comes from the 20 log form used for voltage ratios, which does not apply to power.
C300
Multiplying 10 by 30 mixes up the formula; the decibel value must be divided by 10 first.
D1000 ✓ Correct answer
30 = 10 log(ratio) gives log(ratio) = 3, so the power ratio is 10^3 = 1000.
Reference: OpenStax, College Algebra 2e, Section 6.6 Exponential and Logarithmic Equations; SCFHS Medical Devices – Technician Content Guideline (V1.220901), Mathematics, p. 2
Ask a friend
05Saline (viscosity 1.00×10⁻³ Pa·s) must flow at 0.100 cm³/s through a needle of inner diameter 0.40 mm and length 3.0 cm. Using Poiseuille's law for laminar flow, what pressure difference is needed between the ends of the needle?
Practice question Unit: Physics
AAbout 0.30 kPa
Entering the 0.40 mm diameter as the radius makes r⁴ sixteen times too large, so the pressure comes out sixteen times too small.
BAbout 0.60 kPa
This is one eighth of the true value, which happens when the factor 8 in the numerator of ΔP = 8ηlQ/(πr⁴) is dropped.
CAbout 4.8 kPa ✓ Correct answer
With r = 0.20 mm, Q = 1.0×10⁻⁷ m³/s, ΔP = 8ηlQ/(πr⁴) gives about 4.8×10³ Pa.
DAbout 15 kPa
Leaving π out of the denominator inflates the result by a factor of about 3.14, giving this value.
Reference: OpenStax, College Physics 2e, Section 12.4 Viscosity and Laminar Flow; Poiseuille's Law; SCFHS Medical Devices – Technician Content Guideline (V1.220901), Physics, p. 2
Ask a friend
06A sample of copper contains 60.00% of 63Cu (mass 62.9296 amu) and 40.00% of 65Cu (mass 64.9278 amu). What is the average atomic mass of this sample?
Practice question Unit: Chemistry
A62.93 amu
This is the mass of the lighter isotope alone and ignores the heavier isotope's contribution.
B63.55 amu
This is the average for copper with its natural abundances, not for the 60:40 mixture given here.
C63.73 amu ✓ Correct answer
Weighting each mass by its fraction gives 0.6000 × 62.9296 + 0.4000 × 64.9278 = 63.73 amu.
D63.93 amu
This is the plain mean of the two masses, which would apply only to a 50:50 mixture.
Reference: OpenStax, Chemistry 2e, Section 2.3 Atomic Structure and Symbolism; SCFHS Medical Devices – Technician Content Guideline (V1.220901), Chemistry, p. 2
Ask a friend
07A memory chip has 12 address lines and an 8-bit-wide data bus. What is its total storage capacity in bits?
Practice question Unit: Medical Informatics
A96
Multiplying 12 lines by 8 bits confuses line count with capacity; locations grow as a power of two.
B4,096
Twelve address lines give 4,096 locations, but each holds 8 bits, so 4,096 is a count of bytes, not of bits.
C32,768 ✓ Correct answer
2 to the 12th is 4,096 locations, and 4,096 times 8 bits per location equals 32,768 bits.
D49,152
Multiplying 4,096 locations by 12 uses the number of address lines instead of the data width.
Reference: Kuphaldt, Lessons in Electric Circuits, Vol. IV – Digital (4th ed., 2021), Chapter 15 "Digital Storage (Memory)", Section 15.2 and 15.3; SCFHS Medical Devices – Technician Content Guideline (V1.220901), Medical Informatics, p. 2
Ask a friend
08A technician photographs a failed infusion pump to send to the vendor's support team, and the patient's face and name band are visible. He plans to post the photo on social media to complain about the vendor. What does the Code say about publishing patients' images?
Practice question Unit: Professionalism
APublication is allowed when the patient does not object while being photographed
Not objecting is not written permission, which is what the Code requires for publication.
BWritten permission is needed, and nothing may identify the patient in the image ✓ Correct answer
The Code bars publishing patient images in any media, including social media, unless the patient gives written permission and nothing identifies him.
CPublication is allowed when a small black bar covers only the patient's name band
Hiding only the name band leaves the face visible; the Code requires that nothing identifies the patient.
DPublication is allowed when the purpose is to protect other patients from the device
The Code gives no exception for a good purpose; the rule on written permission and non-identification still applies.
Reference: SCFHS, Code of Ethics for Healthcare Practitioners (2014), Chapter 2 "Healthcare Practitioner’s Duties Towards Patients", Section (F) Photographing Patients and Recording Their Voices, item 7, p. 21; SCFHS Medical Devices – Technician Content Guideline (V1.220901), Professionalism, p. 2
Ask a friend
09Sensor A reads 78 units (population mean 70, SD 5). Sensor B reads 95 units (population mean 80, SD 8). Which statement correctly compares how unusual the two readings are?
Practice question Unit: Research and Design
Az(A) = 1.60 and z(B) = 1.88, so A is further from its mean.
The z-values are right, but 1.60 is smaller than 1.88, so A is the less extreme reading.
Bz(A) = 0.63 and z(B) = 0.53, so A is further from its mean.
These values come from dividing SD by the deviation instead of the deviation by SD, which inverts the ratio.
Cz(A) = 1.60 and z(B) = 1.88, so B is further from its mean. ✓ Correct answer
The z-score (reading − mean) / SD gives 1.60 for A and 1.875 for B, so B is more extreme.
Dz(A) = 8 and z(B) = 15, so B is further from its mean.
Raw differences of 8 and 15 ignore each sensor's spread; they are not z-scores and cannot be compared fairly.
Reference: OpenStax, Introductory Statistics, Section 2.7 Measures of the Spread of the Data (z-scores and comparing values); SCFHS Medical Devices – Technician Content Guideline (V1.220901), Research and Design, p. 2
Ask a friend
10A technician with wet hands touches a metal enclosure that has become energized at 120 V. Assume the wet-body resistance is about 10.0 kΩ and the shock is a 1 s, 60 Hz current through the trunk. Using standard shock-effect values, which effect is expected from the resulting current?
Practice question Unit: Patient Safety, Electrical Safety and Hazards
AThreshold of sensation only, with no other effect on the person
120 V / 10 kΩ is 12 mA, well above the 1 mA threshold of sensation, so sensation is not the only effect.
BVentricular fibrillation is possible, which is often fatal here
Fibrillation needs roughly 100–300 mA through the trunk; 12 mA is about ten times too small for that effect.
CSustained muscular contraction; he may be unable to let go ✓ Correct answer
12 mA falls in the 10–20 mA band, where muscles contract continuously and the hand cannot release the conductor.
DOnset of burns, depending on how concentrated the current is
Burns begin near 300 mA; 12 mA is far below that level, although contact heating could occur at the skin.
Reference: OpenStax, College Physics 2e, Section 20.6 Electric Hazards and the Human Body; SCFHS Medical Devices – Technician Content Guideline (V1.220901), Patient Safety, Electrical Safety and Hazards, p. 2
Ask a friend